How do I troubleshoot jumps and drift in radar flow meter data? What quality-control measures exist?
Most anomalies in radar flow meter data trace back to an interference source or a parameter problem.
Most anomalies in radar flow meter data trace back to an interference source or a parameter problem. Work through this checklist on the HZ-KSLF and about 90% of jump/drift issues resolve on site.
Classify the anomaly first: level jumps usually mean the radar beam is hitting floating debris, weeds or surface foam; velocity jumps come from wind waves, heavy rain striking the surface, or strong EMI; slow drift typically indicates a loosened mounting bracket shifting beam aim, or siltation changing the actual wetted geometry.
Step-by-step checks:
- Inspect the bracket and flange for looseness; verify beam verticality with an inclinometer — deviation should stay under 1°;
- Clear floating debris, cobwebs and bird nests from the beam footprint; keep the antenna face clean;
- Look for newly installed high-power VFDs or motors nearby — they can couple into the EC32 RTU through power lines; add a mains filter and check earthing if needed;
- Verify level against a manual staff gauge; recalibrate the zero point if the difference exceeds 2 cm;
- If velocity reads consistently high or low while level is normal, run a current-meter comparison and re-rate the velocity coefficient.
Quality control: the HZ-KSLF reports a built-in signal quality metric (SNR). On the platform side, enable three-tier QC — single-point range checks (flag physically impossible values), rate-of-change limits (flag suspicious jumps between samples), and level-velocity consistency checks. Combined with raw SNR uploaded via EC32, the platform can distinguish a genuine flood surge from device faults and suppress false alarms.
Set up monthly inspections: bracket tightness, antenna cleanliness, one staff-gauge cross-check — plus a velocity comparison before each flood season.